@inproceedings{1cd4b8d15bc048e387410d39fb754af8,
title = "Miniature ultrasonically powered wireless nerve cuff stimulator",
abstract = "We present a wireless neural stimulator composed of only three discrete components. The capsule was 8 mm long and was designed to be clasped directly upon a peripheral nerve. Power was supplied by low-intensity 1 MHz ultrasound transmitted into the body. The prototype was capable of generating currents in excess of 1 mA. For in vivo testing the device was implanted in a rat hind limb on the sciatic nerve, and when insonated with pulse intensities of 10-150 mW/cm2 the stimulator excited motor axons inducing predictable contractions of the lower leg muscles.",
author = "Larson, {Patrick J.} and Towe, {Bruce C.}",
year = "2011",
month = jul,
day = "20",
doi = "10.1109/NER.2011.5910538",
language = "English (US)",
isbn = "9781424441402",
series = "2011 5th International IEEE/EMBS Conference on Neural Engineering, NER 2011",
pages = "265--268",
booktitle = "2011 5th International IEEE/EMBS Conference on Neural Engineering, NER 2011",
note = "2011 5th International IEEE/EMBS Conference on Neural Engineering, NER 2011 ; Conference date: 27-04-2011 Through 01-05-2011",
}