Abstract
We demonstrate a microfluidic system which applies engineering feedback principles to control the pH of a solution with a high degree of precision. The system utilizes an extended-gate ion-sensitive field-effect transistor (ISFET) along with an integrated pseudo-reference electrode to monitor pH values within a microfluidic reaction chamber. The monitored reaction chamber has an approximate volume of 90 nL. The pH value is controlled by adjusting the flow through two input channels using a pulse-width modulated signal applied to on-chip integrated valves. We demonstrate real-time control of pH through the feedback-controlled stepping of 0.14 pH increments in both the increasing and decreasing direction. The system converges to the pH setpoint within approximately 20 seconds of a step change. The integration of feedback theory into a microfluidic environment is a necessary step for achieving complete control over the microenvironment.
| Original language | English (US) |
|---|---|
| Pages (from-to) | 1191-1197 |
| Number of pages | 7 |
| Journal | Lab on a Chip |
| Volume | 14 |
| Issue number | 6 |
| DOIs | |
| State | Published - Mar 21 2014 |