Development of a system to rapidly measure sampler penetration up to 20 μm aerodynamic diameter in calm air, using the aerodynamic particle sizer

A. D. Maynard, L. C. Kenny, P. E.J. Baldwin

Research output: Contribution to journalArticlepeer-review

16 Scopus citations

Abstract

Although the use of monodisperse aerosols to measure penetration through sampling devices results in accurate measurements, it is time consuming and expensive. Measurement methods using polydisperse aerosols are orders of magnitude faster, but tend to be restricted to smaller particle diameters. This is particularly the case with systems using the Aerodynamic Particle Sizer (APS) in calm air, which are limited by poorly characterised aspiration efficiency and low detection efficiency above approximately 10 μm. By measuring the aspiration of such a system in calm air, and introducing a virtual impactor to act as an aerosol concentrator above 10 μm, we have developed an APS-based penetration measurement system that may be used successfully up to 20 μm. Experimental errors (± 1 S.D.) are typically less than 20-25% at 20 μm, reducing to below an estimated 10% at 15 μm. Errors below 10 μm are of the order of a few percent. The system is ideally suited to the rapid and inexpensive development of size selective aerosol samplers with cut points below 15 μm, and is capable of quantifying the performance of such samplers under calm air conditions.

Original languageEnglish (US)
Pages (from-to)1215-1226
Number of pages12
JournalJournal of Aerosol Science
Volume30
Issue number9
DOIs
StatePublished - Oct 1999
Externally publishedYes

ASJC Scopus subject areas

  • Environmental Engineering
  • Pollution
  • Mechanical Engineering
  • Fluid Flow and Transfer Processes
  • Atmospheric Science

Fingerprint

Dive into the research topics of 'Development of a system to rapidly measure sampler penetration up to 20 μm aerodynamic diameter in calm air, using the aerodynamic particle sizer'. Together they form a unique fingerprint.

Cite this